7.提供軟件開發工具包(Software Development Kit),用戶可自行開發客制或全自動化測試軟件
產品規格
Item
Option
Specifications
Supply voltage
0.5 / 2.828 / 20 / 2 / (70.71) (Vrms)
±1%
CM point voltage
1.414 / 10/ 1/ (35.355) (Vrms)
±1% ( 0.25Vrms ±2% )
Frequency
50 / 60 / 100 /120 (Hz)
±1%
Electrode with Impedance
Change electrode via the touch screen of CMRR 3.0E
None/RA/LA/LL/V1/V2/V3/V4/V5/V6 /ALL
Electrode without Impedance
Change electrode via the touch screen of CMRR 3.0E
RA/LA/LL/V1/V2/V3/V4/V5/V6
Imbalance impedance, R
10k?
10k? ±1%
Imbalance impedance, C
47nF
47nF ± 5%
DC offset
Internal battery powered, can be added on RA/ LA/LL/V1~V6
150mV ± 1% Up to 40hrs with intermittent use
100 pF capacitor
Use 11:1 (110M?:10MΩ) voltage divider to measure indirectly
100pF ± 5%
Environment
Intended for normal laboratory environment. The selection of critical components is known to be stable in the range shown. The 110M? divider may be affected by high humidity in excess of 85%